E040301 Hematology Analyzer
E040301 Hematology analyzer
Technical specification
|
Model |
E040301 |
|
Automatization |
Fully automatic |
|
Throughput |
60 tests/hr |
|
Parameters |
WBC, |
|
RBC,HCT,MCV,RDW-SD,RDW-CV,MCH,MCHC | |
|
PLT,MPV,PCT,PDW | |
|
HGB | |
|
WBC histograms, RBC histograms, PLT histograms | |
|
Measuring principle |
Electro-impedance ,photometric analysis |
|
Sample volume |
10ul(whole blood),20ul(prediluted blood) |
|
Storage |
200.000 results including histograms |
|
Display |
LCD displaying all the reading results of sample |
|
Printer |
Built-in thermal printer or interface to external printer |
|
Data input |
Keypal( mouse and keyboard available) /touch display /key board/barcode |
|
Output |
RS-232,USB,Internet access, available for LIS system |
|
QC |
QC Program calculates X, SD,CV%, and saves data |
|
Language |
English , Chinese , other language on request |
|
Power supply |
AC110/220V 50/60HZ |
|
Dimensions |
|
|
Weight |
|
|
Linearity range |
Accuracy(CV%) | |||
|
WBC |
0.0-99.9(﹡109/L) |
WBC≤2.0% |
RBC≤1.5% |
PCT≤3.0% |
|
RBC |
0.0-99.9(﹡1012/L) |
PLT≤4.0% |
HGB≤1.5% |
MPV≤3.0% |
|
HGB |
0.0-300(g/l) |
HCT≤2.0% |
MCV≤1.0% |
PDW≤3.0% |
|
PLT |
0-999(﹡109/L) |
MCH≤2.0% |
MCHC≤2.0% |
|
Features
|
Easy operation |
Internal real-time monitoring clog system |
|
Carry-over prevented by automatic cleaning probe interior and exterior of probe |
Clog prevention and maintenance function |
|
Automatic standby for 24 hours | |
|
Both whole blood test mode and prediluted blood test mode available |
Multiple report formats available |
|
Free cyanide reagent | |
|
Automatic self-diagnosis and alarm function |
|